Skip to main content

DC defibrillator principle & defibrillator.

                          DC DEFIBRILLATOR:


  •  In almost all present-day DC defibrillators, an energy storage capacitor is charged at a relatively slow rate (seconds) from the AC line by means of a step-up transformer and rectifier arrangement or from a battery and a DC to DC converter arrangement
  • the energy stored in the capacitor is then delivered at relatively rapid rate (millisecond) to the chest of the patient. 
          
  • The basic circuit diagram of a DC defibrillator is shown in Figure.
  • A variable auto-transformer T1 forms the primary of a high voltage transformer T2. The output voltage of the transformer is rectified by a diode rectifier and is connected to a vacuum type high voltage change-over switch.
  • In position A, the switch is connected to one end of an oil-filled 16 µF capacitor. In this position, the capacitor charges to a voltage set by the positioning of the auto- transformer.
  • When the shock is to be delivered to the patient, a foot switch or a push button mounted on the handle of the electrode is operated. The high voltage switch changes over to position ‘B’ and the capacitor is discharged across the heart through the electrodes.
  • The high current electric pulse damages the contractility of the ventricles. This is overcome by inserting a current limiting inductor in series with the patient circuit. The disadvantage of using an inductor is that any practical inductor will have its own resistance and dissipates part of the energy during the discharge process.
  • The inductor also slows down the discharge from the capacitor by the induced counter voltage.
  • The discharge resistance the patient represents is of 50 to 100Ω.
  • A typical discharge pulse of the defibrillator is shown in Figure.
      
  • There are two general classes of waveforms: Mono-phasic and Biphasic.
  • Monophasic waveform use high levels of energy delivered in one direction through the patient’s heart whereas a biphasic waveform delivers energy in both directions. The biphasic waveform is preferred as it defibrillate more effectively than other types of waveforms.
  • for internal defibrillation, energies up to 100 Ws are usually required whereas higher energylevels are necessary for external defibrillation.
  • The DC defibrillator cannot be used for rapidlyrepeated shocks because it requires about 10 s to recharge the capacitor.

 

Comments

Popular posts from this blog

MICROPROCESSOR BASED VENTILATOR

Microprocessor based Ventilator   Figure shows the block diagram of microprocessor controlled ventilator.(given below). It consists of two inter-connected systems: The Pneumatic flow system and an Electronic control system. The pneumatic flow system enables the flow of gas through the ventilator. Oxygen and medical grade air enter the ventilator at 3.5 bar (50 psi) pressure. These gasses enter the air/oxygen mixer where they combine and gasses then enter a large reservoir tank. An electronically controlled flow valve proportions the gas flow from the reservoir tank to the patient breathing circuit.   As the gasses leave the ventilator, they pass by an oxygen analyzer, a safety ambient air inlet valve and a back-up mechanical over pressure valve. The microprocessor controls each valve to deliver the desired inspiratory air and oxygen flows for mandatory and spontaneous ventilation. The electronic control system may use one or...

BED SIDE MONITOR/ MULTI PARA MONITOR

Bed side Monitor/ Multi-para Monitor: The Bed side monitoring system is designed to display an electrocardiogram, heart rate with high and low alarms, pulse rate, dynamic pressure or other waveforms received from external preamplifiers. The system basically consists of three circuit blocks: Preamplifier section, Logic boards and Display part. Various amplified signals are carried to a multiplexer and then to an analog-to-digital converter, included in the logic board. The central processing unit along with memory gives X and Y output for the CRT display. The character generator output is mixed with the Y output for the CRT display. The alarm settings, selection switches for different parameters and the defibrillator synchronization system communicate with the CPU.   The alarm signals are also initiated under its control. The memory comprises 5K bytes ROM and 3.25K bytes RAM with 256 samples of ECG It basically co...