Microprocessor based Ventilator
- Figure shows the block diagram of microprocessor controlled ventilator.(given below).
- It consists of two inter-connected systems: The Pneumatic flow system and an Electronic control system.
- The pneumatic flow system enables the flow of gas through the ventilator.
- Oxygen and medical grade air enter the ventilator at 3.5 bar (50 psi) pressure. These gasses enter the air/oxygen mixer where they combine and gasses then enter a large reservoir tank.
- An electronically controlled flow valve proportions the gas flow from the reservoir tank to the patient breathing circuit.
- As the gasses leave the ventilator, they pass by an oxygen analyzer, a safety ambient air inlet valve and a back-up mechanical over pressure valve.
- The microprocessor controls each valve to deliver the desired inspiratory air and oxygen flows for mandatory and spontaneous ventilation.
- The electronic control system may use one or more microprocessors and software to perform monitoring and control functions in a ventilator.
- These parameters include setting of the respiration rate, flow waveform, tidal volume, oxygen concentration of the delivered breath, peak flow and PEEP.
- The sensors are connected to electronic processing circuits which makes them available for digital readouts. The signals are also compared with pre-set alarm levels so that if they fall outside a pre-determined normal range, alarms are sounded.
- The pressure sensors are normally of semiconductor strain gauge type placed in a bridge configuration.
Block diagram for MICROPROCESSOR BASED VENTILATOR.
Comments
Post a Comment