Skip to main content

CARDIAC MONITOR (type of patient monitoring)

CARDIAC MONITOR
  • The most important physiological parameters monitored in the intensive care unit are heartrate and ECG produced by the heart. This is done to observe the presence of arrhythmias or to detect changes in the heart rate.
  • It is specifically useful for monitoring patients with cardiac problems.
  • It mainly shows the parameters related to heart such as ECG, Heart Rateand Blood Pressure etc.
  • These instruments are also called ‘Cardio scopes’ and comprises of:
             1)     Disposable type pre-gelled electrodes to pick up the ECG signal.
             2)     Amplifier and a CRT/LCD for the amplification and display the ECG waveform
           3)    A heart rate meter to indicate average heart rate with audible beep or flashing light    or                   both     with each beat.                                                                                                                                  4)     An alarm system to produce signal in the event of abnormalities occurring in the heart rate.
 
  • Small cardio scopes using 3” diameter cathode ray tubes are mounted on anesthesia trollies.These are called “Anesthesia monitors’’.
  • These monitors are used by the anesthetist for continuously monitoring of the ECG of anaesthetized patients. 
  • Selection of System Parameters:
  •  In a digital memory system, operating parameters should be carefully selected to maintain signal accuracy, bandwidth.
  •  The following are the important parameters:
1)Sampling rate:Most of the single channel memory display systems have a sampling rate of 250 samples/s and multi-channel display work on 180 samples/s
 
 2)Word Length:The accuracy of analog signalsdepends on the number of bits used in the conversion.
  •  The greater the number of bits per word, the greater will be the resolution.
  • Usually, an 8-bit word is chosen.


3) Memory capacity: Increasing either thesampling rate or the display time interval will increase the memory size proportionally.

 

Comments

Popular posts from this blog

MICROPROCESSOR BASED VENTILATOR

Microprocessor based Ventilator   Figure shows the block diagram of microprocessor controlled ventilator.(given below). It consists of two inter-connected systems: The Pneumatic flow system and an Electronic control system. The pneumatic flow system enables the flow of gas through the ventilator. Oxygen and medical grade air enter the ventilator at 3.5 bar (50 psi) pressure. These gasses enter the air/oxygen mixer where they combine and gasses then enter a large reservoir tank. An electronically controlled flow valve proportions the gas flow from the reservoir tank to the patient breathing circuit.   As the gasses leave the ventilator, they pass by an oxygen analyzer, a safety ambient air inlet valve and a back-up mechanical over pressure valve. The microprocessor controls each valve to deliver the desired inspiratory air and oxygen flows for mandatory and spontaneous ventilation. The electronic control system may use one or...

DC defibrillator principle & defibrillator.

                          DC DEFIBRILLATOR:   In almost all present-day DC defibrillators, an energy storage capacitor is charged at a relatively slow rate (seconds) from the AC line by means of a step-up transformer and rectifier arrangement or from a battery and a DC to DC converter arrangement the energy stored in the capacitor is then delivered at relatively rapid rate (millisecond) to the chest of the patient.             The basic circuit diagram of a DC defibrillator is shown in Figure. A variable auto-transformer T1 forms the primary of a high voltage transformer T2. The output voltage of the transformer is rectified by a diode rectifier and is connected to a vacuum type high voltage change-over switch. In position A, the switch is connected to on...

BED SIDE MONITOR/ MULTI PARA MONITOR

Bed side Monitor/ Multi-para Monitor: The Bed side monitoring system is designed to display an electrocardiogram, heart rate with high and low alarms, pulse rate, dynamic pressure or other waveforms received from external preamplifiers. The system basically consists of three circuit blocks: Preamplifier section, Logic boards and Display part. Various amplified signals are carried to a multiplexer and then to an analog-to-digital converter, included in the logic board. The central processing unit along with memory gives X and Y output for the CRT display. The character generator output is mixed with the Y output for the CRT display. The alarm settings, selection switches for different parameters and the defibrillator synchronization system communicate with the CPU.   The alarm signals are also initiated under its control. The memory comprises 5K bytes ROM and 3.25K bytes RAM with 256 samples of ECG It basically co...